Model SHEATH series probes are internal temperature sensors designed to measure the temperature of liquids, air, and gases. The sheathed section can be bent to a radius five times the outer diameter of the sheath. Specific components may be combined to customize products for the needs of a customer's specific application.
Sleeve type
*Models with an outer sheath diameter of φ0.25 / 0.5mm have sleeve dimensions of φ5 × 30mm and no spring.
Grip type/Fluororesin coating
*Applicable sheath diameter : φ1.0mm(app.φ1.4mm)/φ1.6mm(app.φ2.0mm)/φ3.2mm(app.φ3.6mm)
*Maximum length : 1m
*The demention of "( )" is the outside diameter including the thickness of Fluororesin coating.
*Not manufacturable(J3(Exposed junction))
Sleeve type/Fluororesin coating
*Applicable sheath diameter : φ1.0mm(app.φ1.4mm)/φ1.6mm(app.φ2.0mm)/φ3.2mm(app.φ3.6mm)
*Maximum length : 1m
*Not manufacturable(J3(Exposed junction))
Terminal box type/Fluororesin coating
*Applicable sheath diameter : φ1.0mm(app.φ1.4mm)/φ1.6mm(app.φ2.0mm)/φ3.2mm(app.φ3.6mm), Maximum length : 1m
*Not manufacturable(J3(Exposed junction))
φ0.25mm
*Only (Thermocouple type (K))
*Only (Junction J1(Grounded junction), J2(Ungrounded junction))
*Only (Material M3(Inconel))
φ0.5mm
*Only (Thermocouple type (K))
*Only (Junction J1(Grounded junction), J2(Ungrounded junction))
*Only (Material M3(Inconel))
*The demention of "( )" is the outside diameter including the thickness of Fluororesin coating.
Thermocouple Type | Type E | Type K | |||
Material | SUS316 | SUS316 | SUS310 | Inconel | |
Diameter(mm) | 0.25 | - | - | - | 500 |
0.5 | - | - | - | 500 | |
1.0 | 650 | 650 | 650 | 650 | |
1.6 | 650 | 650 | 650 | 650 | |
3.2 | 750 | 750 | 750 | 750 | |
4.8 | 800 | 800 | 800 | 900 | |
6.4 | 800 | 800 | 800 | 1000 | |
8.0 | 800 | 900 | 900 | 1050 |
*Operating temperature limit : 200°C(Fluororesin coating type)
Diameter | Response time(s) | Standard resistance(Ω/m) | |
Type E | Type K | ||
0.25 | 0.4 | - | 570 |
0.5 | 0.9 | - | 124 |
1.0 | 2 | 38 | 32 |
1.6 | 3 | 15 | 13 |
3.2 | 5 | 5 | 4 |
4.8 | 10 | 2.6 | 2.2 |
6.4 | 15 | 1.2 | 1.0 |
8.0 | 20 | 0.9 | 0.7 |
*Avoid immersing these probes in molten aluminum. Doing so may result in corrosion of the sheathed sensor.
*Make sure that the sum of the sheath and code resistance values does not exceed the value of the signal source resistance of the instrument body.
*Select the same type of thermocouple as thermometer
E
Type E
K
Type K
Thermocouple Type | Temperature range | Tolerance |
Type E | 333°C ≦ t < 900°C | ±0.0075・|t|°C |
-40°C ≦ t < 333°C | ±2.5°C | |
Type K | 333°C ≦ t < 1200°C | ±0.0075・|t|°C |
-40°C ≦ t < 333°C | ±2.5°C |
*t : Measured temperature
Exposed junction
*Not manufacturable (Form (AH,BH,CH))
*Not manufacturable (Diameter(φ0.25,φ0.5))
シース外径(mm) | シース最大長 |
0.25 | L1000(1m)※ |
0.5 | L1000(1m) |
1.0 | L1000(1m) |
1.6 | L1000(1m) |
3.2 | L1000(1m) |
4.8 | L1000(1m) |
6.4 | L1000(1m) |
8.0 | L1000(1m) |
●Max. Length (m)(Please order in increments of 100mm)
Diameter(mm) | Max. Length (m) |
0.25 | L1000(1m)* |
0.5 | L1000(1m) |
1.0 | L1000(1m) |
1.6 | L1000(1m) |
3.2 | L1000(1m) |
4.8 | L1000(1m) |
6.4 | L1000(1m) |
8.0 | L1000(1m) |
*The length of J2(ungrounded type) is up to 0.5m.
TS1
Diameter less than φ0.5mm
Standard specifications : TS1m
Outer diameter : φ2.3mm
Sheath material : Silicone rubber
Heat resistant : 240°C
TC1
Diameter more than φ1.0mm
Standard specifications : TC1m
Outer diameter : φ4mm
Sheath material : Silicone rubber
Heat resistant : 240°C
The cable length can be specified in 0.5m increments.
Anritsu standard plug
For HR-1*0*,AP-400 etc.
Miniature plug
for AM-8***,HR-1*5*,AP-450,AR-650* etc.
Without plug
Spade terminal for φ3mm, φ4mm, φ5mm
Ring terminal for φ3mm, φ4mm, φ5mm
Spade terminal for φ3mm, φ4mm (made of thermocouple)
Ring terminal for φ3mm, φ4mm (made of thermocouple)
Sheathed banana plug