●Tolerance
Thermocouple typeTemp. rangeClassTolerance
Type E333°C ≦ t < 900°CClass2±0.0075・|t|°C
-40°C ≦ t < 333°C±2.5°C

Type K

333°C ≦ t < 1200°C±0.0075・|t|°C
-40°C ≦ t < 333°C±2.5°C

* t=temperature

● Response time and Standard resistance
Outside diameter(mm)Response time(s)Standard resistance(Ω/m)
Type EType K
0.250.4-570
0.50.9-124
1.023832
1.631513
3.2554
4.8102.62.2
6.4151.21.0
8.0200.90.7

*Avoid immersing these probes in molten aluminum. Doing so may result in corrosion of the sheathed sensor.

*Make sure that the sum of the sheath and code resistance values does not exceed the value of the signal source resistance of the instrument body.

●Max. Temp.(°C)
Thermocouple typeType EType K
MaterialSUS316SUS316SUS310SInconel

Diameter

(mm)

0.25---500
0.5---500
1.0650650650650
1.6650650650650
3.2750750750750
4.8800800800900
6.48008008001000
8.08009009001050

*Operating temperature limit 200°C with Fluororesin coating.