High Superior Surface Probes(Contact support type) Model SX series
Type E Model:SX-371E-01-1-TPC1-ANP
Type K Model:SX-371K-01-1-TPC1-ANP

画像

Back

Product Info

Temp. range
-50 to 800°C
Tolerance
-50°C ≦ t < 333°C : ±2.5°C
333°C ≦ t ≦ 500°C : ±(0.0075 × | t |)°C
500°C < t ≦ 700°C : ±(0.015 × | t |)°C
700°C < t ≦ 800°C : ±(0.02 × | t |)°C
* t : Measured temperature
Response time
1 s
Durability
More than 60,000 contacts (at 800°C surface)
Calibration points of test report
100 , 300 , 500°C
allowable tilt
±4°
Cable type and length
Type E : TPC-E 1m
Type K : TPC-K 1m
Plug
Type E : ANP-E-M-L
Type K : ANP-K-M-L
Repair
Repairable

カスタマイズする

Remarks

* The operating temperature limit is determined by the allowable temperature limit of the probe head contacts the measurement target.
Note that the operating temperature limit is not the same as the allowable temperature limits of the grip,cable,and plug.
* The response time is the time required to detect 99% of the true value on a flat and smooth metal surface.
* Durability is the number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface.
* Please contact the distributor for price and delivery.