Q3 What types of thermocouples are available?

The following table lists typical thermocouples.
Of them, Anritsu Meter employs two types E and K as standard ones because they deteriorate less than the types J and T in an oxidizing atmosphere and are marketable at lower prices than the types R, B and S.
With a wider measurable temperature range, they are applicable to a variety of uses.

 

TypeAlloy combination
EChromel-Constantan
KChromel-Alumel
JIron-Constantan
TCopper-Constantan
RPlatinum (13% rhodium)-Platinum
BPlatinum (30% rhodium)-Platinum (6% rhodium)
SPlatinum (10% rhodium)-Platinum

 

Type Composing materialsTolerance
KChromel(+)
Alumel(-)
Class 1-40°C ≦ t <375°C : ±1.5°C
375°C ≦ t <1000°C : ±0.004・|t|°C
Class 2-40°C ≦ t <333°C:±2.5°C
333 °C ≦ t <1200 °C : ±0.0075・|t|°C
Class 3-167 °C ≦ t <40 °C : ±2.5 °C
-200 °C ≦ t <-167 °C : ±0.015・|t|°C
EChromel(+)
Constantan(-)
Class 1-40°C ≦ t <375°C : ±1.5°C
375°C ≦ t <800°C : ±0.004・|t|°C
Class 2-40°C ≦ t <333°C : ±2.5°C
333 °C ≦ t <900 °C : ±0.0075・|t|°C
Class 3-167 °C ≦ t <40 °C : ±2.5 °C
-200 °C ≦ t <-167 °C : ±0.015・|t|°C
JIron(+)
Constantan(-)
Class 1-40°C ≦ t <375°C : ±1.5°C
375°C ≦ t <750°C : ±0.004・|t|°C
Class 2-40°C ≦ t <333°C : ±2.5°C
333 °C ≦ t <750 °C : ±0.0075・|t|°C
Class 3
TCopper(+)
Constantan(-)
Class 1-40°C ≦ t <125°C : ±0.5°C
125°C ≦ t <350°C : ±0.004・|t|°C
Class 2-40°C ≦ t <133°C : ±1°C
133 °C ≦ t <350 °C : ±0.0075・|t|°C
Class 3-67 °C ≦ t <40 °C : ±1 °C
-200 °C ≦ t <-67 °C : ±0.015・|t|°C
RPlatinum/13%Rhodium(+)
Pure Platinum(-)
Class 10°C ≦ t <1100°C : ±1°C
1100°C ≦ t <1600°C : ±[1°C+0.003・(|t|-1100°C)]°C
Class 20°C ≦ t <600°C : ±1.5°C
600 °C ≦ t <1600 °C : ±0.0025・|t|°C
Class 3
BPlatinum/30%Rhodium(+)
Platinum/6%Rhodium(-)
Class 1
Class 2600 °C ≦ t <1700 °C : ±0.0025・|t|°C
Class 3600°C ≦ t <800°C : ±4°C
800 °C ≦ t <1700 °C : ±0.005・|t|°C
SPlatinum/13%Rhodium(+)
Pure Platinum(-)
Class 10°C ≦ t <1100°C : ±1°C
1100°C ≦ t <1600°C : ±[1°C+0.003・(|t|-1100°C)]°C
Class 20°C ≦ t <600°C : ±1.5°C
600 °C ≦ t <1600 °C : ±0.0025・|t|°C
Class 3

※|t| is a value displayed in temperature (°C) regardless of the + or - sign of the measured temperature.