In the case of the internal temperature sensor, apply it to the steam of boiled water from a kettle, etc. to check if 100°C or closer value is displayed.
In the case of the surface temperature sensor, check the thin plate, called a contact plate, of the temperature sensing part at the nose for deformation or any adhered substances.