Measurement Range:K -200 to 1370°C / E -200 to 800°C
(Temperature probes not included)
Model | HR-1300 | ||
Display | LCD | ||
Waterproof specifications | - | ||
Input connector | ASP | ||
Input type | Thermocouple Input: Type E, K | ||
Channel | 1ch | ||
Signal source resistance | MAX.1k ohm | ||
Measurement Range | 1°C Resolution | E | -200 to 800°C |
K | -200 to 1370°C | ||
0.1°C Resolution | E | -104.9 to 504.9°C (Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | |
K | -104.9 to 504.9°C (Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | ||
Measurement Accuracy | 1°C Resolution | 0°C to | ± (0.1% of reading +1°C) |
to 0°C | ± (0.5% of reading +1°C) | ||
0.1°C Resolution | 0°C to | ± (0.05% of reading + 0.2°C) | |
to 0°C | ± (0.15% of reading + 0.2°C) | ||
Reference junction compensation accuracy | ±0.2°C at 25°C±10°C | ||
Temperature Coefficient*1 | ±0.02×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.3°C is added at room temp. of 0°C or 50°C) | ||
Operation Environment | 0 to 50°C, 0 to 80%RH (Non condensing) | ||
Storage Environment | -20 to 50°C, 0 to 85%RH (Non condensing) | ||
Battery life | approx.600h | ||
Power supply(Dry battery) | 4 AA Alkaline batteries | ||
Power supply(AC power) option | available | ||
Sampling rate | approx.200ms | ||
Linearlizer | Digital Linearizer method (Compliant with JIS C1602-2015) | ||
Dimensions(mm) | approx. 82(w)×166(H)×36(D) [Protruded parts excluded] | ||
Weight | approx. 350g [Include Batteries] | ||
Supplied accessories | Manual,Test report,Soft case, | ||
Hand strap,4 AA Alkaline batteries | |||
Optional accessories | Alarm output cable | ||
Alarm output | TABLE1 | ||
Compatible standards | EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial) RoHS:IEC EN 63000:2018 |
*1 Only when the temp. exceeds 25°C±10°C
※ Details of typeJ,T,R(option)
Output Pattern | Upper Limit | Lower Limit |
More than upper limit | Close | Open |
Between upper and lower limit | Open | Open |
Less than lower limit | Open | Close |
Alarm output cable | ANGHA-1.5(accessory) |
※ The alarm output uses a photo MOS relay. ON resistance(internal protection resistance 400Ω, photo MOS relay ON resistance 50Ω) Drive voltage MAX.25V
model : AD-115-500-HR-R (for AC115V)
model : AD-220-500-HR-R (for AC220V)
for HR-1300,HR-1350,HR-1301,HR-1351,HR-1500,HR-1550
* Not CE-approval
model : SCHR
This product is available for repair. Please ask our dealer.