Thermologger AM-9100E / 9100K

サーモロガー AM-9000E / 9000K

Measurement Range
E -200 to 800°C / K -200 to 1370°C

Channel
12ch (ANP connector (same type thermocouple))

Input Type
Thermocouple Input : Type E, K (typeJ,T,R(option))
Voltage input

Interface
USB2.0(microーB)

Interval
(The following numbers are time of interval that can be set.)
less than 1s : 0.2s ~ 0.99s Over 1s : 1s ~ 99m 59s

(Temperature probes not included)

Download catalog

Specifications

ModelAM-9100
Number of inputs12ch : ANP connector (Miniature Plug)
Input connectorANP connector (same type thermocouple)
Input typeTemperature: E, K, J, T, R by selected model *2
DC Voltage:V
Measurement rangeTemperature: TABLE Ⅰ  DC Voltage : ±20.000V
ResolutionTemperature: 0.1℃/1℃  DC Voltage : 0.001V
Measurement accuracyTemperature: TABLE Ⅱ  DC Voltage : ±(0.05% of reading +0.003V)
Memory capacity120,000 data
InterfaceUSB2.0 (micro-B)

Dimensions
*Protruded parts excluded

96(W)×160(H)×60(D)
Weight(Including dry batteries)approx.480g
Reference junction
compensation accuracy
±0.3℃*3
Temperature coefficient± (0.04 × Δt ℃) *4
Time interval

0.2 second to 99 minutes 59 seconds

Signal source resistance500Ω or less (measurement interval 1 second or more)/100Ω or less (measurement interval less than 1 second)
Safe excitation voltageBetween input pins on same channels :±30VDC,Between input pins on adjacent channels : 150Vp-p
LinearlizerDigital Linealizer method (Compliant with JIS C1602-2015)
DisplayGraphic liquid crystal display, 128×64 dots
Environment operational
Environment storage

Operation : 0〜40℃、15〜80%RH (Non condensing)  
Storage : -10〜50℃、10〜85%RH (Non condensing)

Battery lifeTABLE Ⅲ
Power supply4 AA Alkaline batteries(Supplied accessories) or AC-Adaptor(Optional accessories)
Supplied accessoriesManual, Test report, Softcase, Communications cable, 4 AA Alkaline batteries
Optional accessoriesSpecial data analysis software, AC-Adaptor, Carrying case
Compatible standardsCE Marking, RoHS2

The selected thermocouple type (E・K・J・T・R) is entered in ※ of the model name.
*1 Thermocouple type (E・K・J・T・R)
*2 Type J・T・R are optional
*3 At 25℃±10℃
*4 When the ambient temperature is other than from 15℃ to 35℃, "the excess temperature from the ambient temperature Δt℃" x 0.04 is added to the measurement accuracy.
 

TABLE Ⅰ : Measurement Range

Thermocouple typeMeasurement range
E-200 to 800°C
K-200 to 1370°C
J-200 to 1100°C
T-200 to 400°C
R0 to 1760°C

 

TABLE Ⅱ : Measurement accuracy

In the environment of 25 ℃ ± 10 ℃, when the measurement interval is 1 second or more without including the reference junction compensation accuracy.

Thermocouple typeAccuracy compensation rangeMeasurement accuracy
E,K,J,T-200.0~-100.1℃±1.0℃
-100.0~-0.1℃±0.5℃
0.0~200.0℃±0.3℃
200.1~500.0℃±0.5℃
500.1℃ or more±0.7℃
R0.0℃~500.0℃±1.5℃
500.1℃ or more±2.0℃

*If the measurement interval is less than 1 sec, add ± 0.5 ℃ at type E ・ K ・ J ・ T and add ± 1.0℃ at type R

TABLEⅢ 12ch model standard time for continuous use (at 25 ℃, using alkaline AA batteries)

Setting interval0.2sec1sec10sec
AM-9000 Seriesapprox. 50 hoursapprox. 55 hoursapprox. 140 hours

* If the number of used channels is less than 12, the usable time will be longer.

Optional accessories

画像

AC adapter

AD-115-500-HR-R (for AC115V)
AD-220-500-HR-R (for AC220V)

画像

Carrying case

AM-CC2

About repair

This product is available for repair. Please ask our dealer.