Temperature Calibration System ACSⅡ-2000

画像

For Surface Temperature Max. 500°C

Surface Range:30 to 500°C

Download catalog

Overall Accuracy

ModelACSII-2000
Calibration ObjectSurface
Range30.0 to 100.0°C+0.5°C
-0.7°C
100.1 to 200.0°C+0.5°C
-1.1°C
200.1 to 300.0°C+0.5°C
-1.5°C
300.1 to 400.0°C+0.4°C
-2.0°C
400.1 to 500.0°C-0.1°C
-2.9°C

 

 

Control Unit

ModelACSII-2000
Range30 to 500°C
Control PrinciplePID Auto-Tuning Control
ControlInputPt100Ω
Accuracy±0.2% of reading+0.3°C
Display8mm high LED
power supplyAC100V~240V *
Power Consumptionapp.1100W
Dimensions(W×H×D)320*177*350mm
Weight4.5kg

* Please select at purchase
 

Calibration Unit

ModelACSII-2000
Surface±0.1°C
Heater1000W
Overheating prevention mechanismBimetal breaker
Heating time30 → 250°C 30min
30 → 500°C 60min
Cooling time500 → 250°C 40min
250 → 50°C 90min
Dimensions(W×H×D)390*221*360mm
Weight20kg


Supplied accessories

ModelACSII-2000
Supplied accessories

Power cable,Heater cable, Sensor cable, Manual,Test report,

Protective cover for calibration block



About repair