Temperature Calibration System ACSⅡ-3000

画像

For Surface Temperature Max. 500°C

Internal Range:30 to 500°C

Download catalog

Overall Accuracy

ModelACSII-3000
Calibration ObjectInternal
Range30.0 to 100.0°C±0.26°C
100.1 to 200.0°C±0.32°C
200.1 to 300.0°C±0.37°C
300.1 to 400.0°C±0.43°C
400.1 to 500.0°C±0.48°C


Control Unit

ModelACSII-3000
Range 30 to 500℃
Control PrinciplePID auto-Tuning Control
ControlInputPt100Ω
Accuracy±0.2% of reading+0.3°C
Display8mm high LED display
Reference sensorInputPt100Ω Φ4.8mm JIS C 1604
Accuracy±(0.05% of reading+0.05)°C
Display12.5mm high LED display
Power supplyAC100V~240V *
Power Consumption1600W
Dimensions(W×H×D)430*177*350mm
Weight5kg

* Please select at purchase
 

Calibration Unit

ModelACSII-3000
Point8 holes based on specified dimension
Surface±0.1℃
Heater1500W
Overheating prevention mechanismBimetal breaker
Heating time30 → 250°C 60min
30 → 500°C 150min
Cooling time500 → 250°C 220min
250 → 100°C 450min
Dimensions(W×H×D)450*405*460mm
Weight70kg


Supplied accessories

ModelACSII-3000
Supplied accessories

Power cable,Heater cable, Sensor cable, Manual,Test report,

Protective cover for calibration block,

Standard sensor for calibration(Pt 100Ω platinum resistor)



About repair