Model | ACSII-1000 | |
Calibration Object | Surface | |
Range | 30.0 to 100.0°C | +0.5℃ |
-0.7℃ | ||
100.1 to 200.0°C | +0.5℃ | |
-1.1℃ | ||
200.1 to 300.0°C | +0.5℃ | |
-1.5℃ |
Model | ACSII-1000 | |
Range | 30 to 300°C | |
Control Principle | PID auto-Tuning Control | |
Control | Input | Pt100Ω |
Accuracy | ±0.2% of reading+0.3°C | |
Display | 8mm high LED display | |
Power supply | AC100V~240V * | |
Power Consumption | app.250W | |
Dimensions(W×H×D) | 320*177*350mm | |
Weight | 4.5kg | |
* Please select at purchase
Model | ACSII-1000 |
Surface | ±0.1℃ |
Heater | 200W |
Overheating prevention mechanism | Bimetal breaker |
Heating time | 30 → 150°C 25min |
30 → 300°C 50min | |
Cooling time | 300 → 150°C 80min |
150 → 50°C 120min | |
Dimensions(W×H×D) | 260*130*260mm |
Weight | 7kg |
Model | ACSII-1000 |
Supplied accessories | Power cable,Heater cable, Sensor cable, Manual,Test report, Protective cover for calibration block |
This product is available for repair. Please ask our dealer.