Specifications of TH Series / Flat Head Probes

Model number*1TH-11TH-14
Thermocouple typeType E or K
Temp. range*2-50 to 400°C-50 to 500°C
Tolerance*30°C±2.5°C
100°C±2.5°C
200°C±2.5°C
300°C±2.5°C
400°C±3.0°C
500°C-±3.8°C
Tolerance calculation method
t:Temperature(°C)
At over 0°C within temperature range
±2.5°C or ± (0.0075 × | t |)°C, whichever is greater
Response time*42s
Durability*5A(More than 50,000 contacts)

Grip material

Phenol ResinPolyacetal
Repair*6Repairable

*1 The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is available at 0°C or above within the operating temperature on a stationary flat and smooth metal surface. For tolerance at temperature not listed in the above table, please refer to calculation method of tolerance.
*4 The response time is the time required to detect 99% of the true value on a flat and smooth metal surface.
*5 Number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface at a temperature of 300°C.
*6 Please refer to page repair.