Specifications of Pipe inside probes

Model number*1model 305model 344
Thermocouple typeType E or K
Temp. range*2-50 to 220°C-50 to 400°C
Tolerance*30°C ≦ t ≦ 100°C±2.5°C±2.5°C
100°C < t ≦ 200°C±2.5°C±4.4°C
200°C < t ≦ 300°C±2.8°C(below 220°C)±6.3°C
300°C < t ≦ 400°C-±8.1°C
Response time*42s5s
Durability*5More than 5,000 contacts
Pipe materialStainless(SUS304)
Grip materialPhenol Resin
Repair*6Non-repairableRepairable

*1 The model number after thermocouple type is omitted.
*2 The operating temperature limit is determined by the allowable temperature limit of the sensor head contacts the measurement target. Note that the operatingtemperature limit is not the same as the allowable temperature limits of the grip, cable, and plug.
*3 Tolerance is available at 0°C or above within the operating temperature on a stationary flat and smooth metal surface.
*4 The response time is the time required to detect 99% of the true value on a flat and smooth metal surface.
*5 Number of contacts enabling measurement within the tolerance range on a flat and smooth metal surface at a temperature of 300°C (or at the operating temperature limit if the operating temperature limit is below 300°C)
*6 Please refer to page repair.