Model | HR-1100 | HR-1200 HR-1250 HR-1750 | HR-1300 HR-1350 | HR-1301 HR-1351 | HR-1400 HR-1450 | HR-1500 HR-1550 | ||
Display | LCD | LED | LCD | |||||
Waterproof specifications/Equivalent to IPX5 | 〇 | 〇 | - | - | 〇 | - | ||
Input connector | ASP model:HR-1*0* /ANP model:HR-1*5* | |||||||
Input type | Thermocouple Input: Type E, K (Not switchable) | |||||||
Channel | 1ch | |||||||
Signal source resistance | MAX.1 Kohm | |||||||
Measurement Range | 1°C Resolution | E | -200~800°C | |||||
K | -200~1370°C | |||||||
0.1°C Resolution | E | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | ||||||
K | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | |||||||
Measurement Accuracy | 1°C Resolution | 0°C~ | ±(0.1% of reading +1°C) | |||||
~0°C | ±(0.5% of reading +1°C) | |||||||
0.1°C Resolution | 0°C~ | ±(0.05% of reading + 0.2°C) | ||||||
~0°C | ±(0.15% of reading + 0.2°C) | |||||||
Reference junction compensation accuracy | ±0.2°C at 25°C±10°C | |||||||
Temperature Coefficient*1 | ±0.02×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.3°C is added at room temp. of 0°C or 50°C) | |||||||
Operation Environment | 0 to 50°C, 0 to 80%RH (Non condensing) | |||||||
Storage Environment | -20 to 50°C, 0 to 85%RH (Non condensing) | |||||||
Battery life | approx.900h | approx.900h | approx.600h | approx.20h*2 | approx.300h | approx.550h | ||
Power supply(Dry battery) | 4 AA Alkaline batteries | |||||||
Power supply(AC power) | - | - | available | - | available | |||
Sampling rate | approx.200ms | |||||||
Linearlizer | Digital Linearizer method (Compliant with IEC 60584-1 (2013)) | |||||||
Dimensions(mm) | approx. 82(w)×166(H)×36(D) [Protruded parts excluded] | |||||||
Weight | approx. 350g [Include Batteries] | |||||||
Supplied accessories | Manual,Test report,Soft case, Hand strap,4 AA Alkaline batteries | |||||||
Optional accessories | - | - | Alarm output cable | - | Communication cable,software | |||
Analog output cable | - | - | ||||||
Alarm Output | - | - | 〇TABLE1 | 〇TABLE1 | - | ‐ | ||
Analog Output | - | - | - | 〇TABLE2 | ‐ | ‐ | ||
Memory function | - | - | - | - | - | 〇TABLE3 | ||
Compatible standards(CE) | EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial) RoHS:IEC EN 63000:2018 |
*1 Only when the temp. exceeds 25°C ±10°C
*2 Approx. 400h when the analog output is switched always off.
Output Pattern | Upper Limit | Lower Limit |
More than upper limit | Close | Open |
Between upper and lower limit | Open | Open |
Less than lower limit | Open | Close |
Alarm output cable | ANGHA-1.5(accessory) |
※ The alarm output uses a photo MOS relay. ON resistance(internal protection resistance 400Ω, photo MOS relay ON resistance 50Ω) Drive voltage MAX.25V
Rate | 1mV/°C(1°C resolution) | 10mV/°C(0.1°Cresolution) | |
Range | The entire measurement range with 1°C resolution | The entire measurement range with 0.1°C resolution | |
Analog output conversion accuracy(under 25°C±10°C) | Equivalent to the value obtained by adding ±1°C(1mV)to the measurement accuracy | Equivalent to the value obtained by adding ±0.1°C(1mV)to the measurement accuracy | |
Temperature conficient*1 | ±(0.1mV/°C×Δt°C) e.g.±1.5mV(2°C) is added at room temp.of 0°C or 50°C | ±(0.1mV/°C×Δt°C) e.g.±1.5mV(0.2°C) is added at room temp.of 0°C or 50°C | |
When the probe is disconnected | Approx.-2.3V | ||
Measurement range+over | Approx.5.1V | ||
Measurement range-over | Approx.-2.2V | ||
Insulation resistance | 100MΩ/DC500V(between input and analog output) | ||
Withstand voltage | 300Vpーp(between input and analog output) | ||
Cable | ANGHA-1.5(accessory) |
*1 When the temperature exceeds 25°C±10°C ,multiply the excess temperature Δt°C by a coefficient and add it to the analog output conversion accuracy.
※Regarding analog output: The displayed value is D/A converted and output.The update timing is about 200ms, and the output resolution isin 1mV units.When not using the analog output, turn off the analog output.
Time interval | second:1, 5, 10, 30 minutes:1, 5, 10, 30, 60 and manual memory |
Memory capacity | 19999data |