Handheld Thermometer Model HD-1000 series

List of specifications

ModelHD-1100HD-1200HD-1201HD-1202HD-1300HD-1301HD-1302HD-1400HD-1500
DisplayLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLEDLCD with Backlight
Waterproof specificationsIPX5IPX5-----IPX5-
Input connectorASP
Input typeThermocouple Input: Type E, K (Not switchable)
Channel1ch
Signal source resistanceMAX.500ohm
Measurement Range1°C ResolutionE-200~800°C
K-200~1370°C
0.1°C ResolutionE-104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
K-104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
Measurement Accuracy1°C Resolution0°C~±(0.1% of reading +1°C)
~0°C±(0.5% of reading +1°C)
0.1°C Resolution0°C~±(0.05% of reading + 0.2°C)
~0°C±0.5°C
Reference junction compensation accuracy±0.2°C at 25°C±10°C 
Temperature Coefficient±0.01% of F/S per °C (F/S : full scale)
Operation Environment0 to 40°C, 0 to 80%RH (Non condensing)
32 to 104°F, 0 to 80%RH (Non condensing)
Storage Environment-20 to 50°C, 0 to 85%RH (Non condensing)
-4 to 122°F, 0 to 85%RH (Non condensing)
Battery life300h300h150h150h300h150h150h150h150h
Power supply(Dry battery)4 AA Alkali Batteries
Power supply(AC power)--

AC-Adaptor (Option)

-

AC-Adaptor

(Option)

Sampling rateapprox.300ms

approx.

300ms

*1

LinearlizerDigital Linearizer method (Compliant with IEC 60584-1 (2013))
Dimensions(mm)approx. 76(w)×167(H)×36(D)  [Protruded parts excluded]
Weightapprox. 350g [Include Batteries]
Supplied accessoriesManual,Test report,Soft case, Hand strap,4 AA Alkaline batteries
Optional accessories--Analog output CableAlarm output Cable-

Communi

cation

Cable

-Analog output Cable

Software

(AMS-100)

Analog Output--

1mV/°C

(table1)

10mV/°C

(table1)

-

1mV/°C

(table1)

10mV/°C

(table1)

--
Alarm Output----

(table2)

(table2)

(table2)

--
Memory function--------

(table3)

Compatible standardsCE Marking, RoHS

*1 Sampling period will be a user-selected time interval when time interval is specified at 10 seconds or more.

*2 Details of typeJ,T,R

Model

HD-1150

HD-1650

HD-1250

HD-1750

HD-1251HD-1252HD-1350HD-1351HD-1352HD-1450HD-1550
DisplayLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLCD with BacklightLEDLCD with Backlight
Waterproof specificationsIPX5IPX5-----IPX5-
Input connectorANP
Input typeThermocouple Input: Type E, K (Not switchable)
Channel1ch
Signal source resistanceMAX.500ohm
Measurement Range1°C ResolutionE-200~800°C
K-200~1370°C
0.1°C ResolutionE-104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
K-104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
Measurement Accuracy1°C Resolution0°C~±(0.1% of reading +1°C)
~0°C±(0.5% of reading +1°C)
0.1°C Resolution0°C~±(0.05% of reading + 0.2°C)
~0°C±0.5°C
Reference junction compensation accuracy±0.2°C at 25°C±10°C 
Temperature Coefficient±0.01% of F/S per °C (F/S : full scale)
Operation Environment0 to 40°C, 0 to 80%RH (Non condensing)
32 to 104°F, 0 to 80%RH (Non condensing)
Storage Environment-20 to 50°C, 0 to 85%RH (Non condensing)
-4 to 122°F, 0 to 85%RH (Non condensing)
Battery life300h300h150h150h300h150h150h150h150h
Power supply(Dry battery)4 AA Alkali Batteries
Power supply(AC power)--

AC-Adaptor (Option)

-

AC-Adaptor

(Option)

Sampling rateapprox.300ms

approx.

300ms

*1

LinearlizerDigital Linearizer method (Compliant with IEC 60584-1 (2013))
Dimensions(mm)approx. 76(w)×167(H)×36(D)  [Protruded parts excluded]
Weightapprox. 350g [Include Batteries]
Supplied accessoriesManual,Test report,Soft case, Hand strap,4 AA Alkaline batteries
Optional accessories--Analog output CableAlarm output Cable-

Communi

cation

Cable

-Analog output Cable

Software

(AMS-100)

Analog Output--

1mV/°C

(table1)

10mV/°C

(table1)

-

1mV/°C

(table1)

10mV/°C

(table1)

--
Alarm Output----

(table2)

(table2)

(table2)

--
Memory function--------

(table3)

Compatible standardsCE Marking, RoHS

*1 Sampling period will be a user-selected time interval when time interval is specified at 10 seconds or more.

*2 Details of typeJ,T,R

Table1 Analog Output (HD-1201,HD-1202,HD-1301,HD-1302,HD-1251,HD-1252,HD-1351,HD-1352)

 

HD-1201

HD-1251

HD-1301

HD-1351

HD-1202

HD-1252

HD-1302

HD-1352

Rate1mV/°C10mV/°C
RangeAll range at 1°C resolutionAll range at 0.1°C resolution
Resistance100ohm
Accuracy± (0.15% of F/S + 1mV) at 25°C±5°C± (0.15% of F/S + 1mV)
±(0.01% of F/S) except 25°C±5°C


Table2 Alarm Output (HD-1300,HD-1301,HD-1302,HD-1350,HD-1351,HD-1352)

 Output PatternUpper LimitLower Limit
More than upper limitH(5V)L(0V)
Between upper and lower limitL(0V)L(0V)
Less than lower limitL(0V)H(5V)

* Signal current of up to 5mA can be taken out.

Table3 Memory function (HD-1500,HD-1550)

Time intervalsecond:1, 5, 10, 30 minutes: 1, 5, 10, 30, 60 and manual memory
Memory capacity9999data

 

Close the window